N.H. Admin. Code § Saf-C 6802.03

Current through Register No. 50, December 12, 2024
Section Saf-C 6802.03 - Laboratory Evaluation
(a) The laboratory shall evaluate each manufacturer make and model of preliminary breath test device submitted in accordance with Saf-C 6802.02.
(b) The evaluation shall include the following categories:
(1) Analysis of a series of alcohol samples at known concentrations, to verify the accuracy and precision of the approved instrument as meeting the following criteria:
a. For known alcohol concentrations greater than 0.000 g/210L but less than or equal to 0.050 g/210L:
1. The accuracy of the readings shall deviate no more than +/- 0.003 g/210L from known alcohol concentrations; and
2. The total precision range of the readings shall not exceed 0.005 g/210L;
b. For known alcohol concentrations greater than 0.050 g/210L but less than or equal to 0.100 g/210L:
1. The accuracy of the readings shall deviate no more than +/- 0.005 g/210L from known concentrations; and
2. The total precision range of the readings shall not exceed 0.005 g/210L; and
c. For known alcohol concentrations greater than 0.100 g/210L:
1. The accuracy of the readings shall deviate no more than +/- 5% from known alcohol concentrations; and
2. The total precision range of the readings shall not exceed 5% of the target value;
(2) Ease of use in handling the device while performing a test;
(3) Performance under normal and adverse ambient conditions; and
(4) Effects of other substances or sources of energy that might interfere with the proper functioning of the device.
(c) Each device evaluated in accordance with Saf-C 6802.03 shall meet or exceed the performance statements provided by the manufacturer.

N.H. Admin. Code § Saf-C 6802.03

(See Revision Note at chapter heading for Saf-C 6800) #9685, eff 3-23-10

Amended by Volume XXXVIII Number 19, Filed May 10, 2018, Proposed by #12518, Effective 4/23/2018, Expires4/23/2028.