Current through Register No. 50, December 12, 2024
Section Saf-C 6802.03 - Laboratory Evaluation(a) The laboratory shall evaluate each manufacturer make and model of preliminary breath test device submitted in accordance with Saf-C 6802.02.(b) The evaluation shall include the following categories: (1) Analysis of a series of alcohol samples at known concentrations, to verify the accuracy and precision of the approved instrument as meeting the following criteria: a. For known alcohol concentrations greater than 0.000 g/210L but less than or equal to 0.050 g/210L: 1. The accuracy of the readings shall deviate no more than +/- 0.003 g/210L from known alcohol concentrations; and2. The total precision range of the readings shall not exceed 0.005 g/210L;b. For known alcohol concentrations greater than 0.050 g/210L but less than or equal to 0.100 g/210L: 1. The accuracy of the readings shall deviate no more than +/- 0.005 g/210L from known concentrations; and2. The total precision range of the readings shall not exceed 0.005 g/210L; andc. For known alcohol concentrations greater than 0.100 g/210L: 1. The accuracy of the readings shall deviate no more than +/- 5% from known alcohol concentrations; and2. The total precision range of the readings shall not exceed 5% of the target value;(2) Ease of use in handling the device while performing a test;(3) Performance under normal and adverse ambient conditions; and(4) Effects of other substances or sources of energy that might interfere with the proper functioning of the device.(c) Each device evaluated in accordance with Saf-C 6802.03 shall meet or exceed the performance statements provided by the manufacturer.N.H. Admin. Code § Saf-C 6802.03
(See Revision Note at chapter heading for Saf-C 6800) #9685, eff 3-23-10
Amended by Volume XXXVIII Number 19, Filed May 10, 2018, Proposed by #12518, Effective 4/23/2018, Expires4/23/2028.