A licensee authorized to use a remote afterloader unit for medical use must perform full calibration measurements on each unit:
To satisfy subpart 1, full calibration measurements must include, as applicable, determination of:
A licensee must use the dosimetry system described in part 4731.4468, subpart 1, to measure the output.
A licensee must make full calibration measurements required under subpart 1 according to published protocols accepted by nationally recognized bodies.
In addition to the requirements for full calibrations for low dose-rate remote afterloader units under subpart 2, a licensee must perform an autoradiograph of the source to verify inventory and source arrangement at intervals not exceeding one quarter.
For low dose-rate remote afterloader units, a licensee may use measurements provided by the source manufacturer that are made according to subparts 1 to 5.
A licensee must mathematically correct the outputs determined in subpart 2, item A, for physical decay at intervals consistent with one percent physical decay.
Full calibration measurements required under subpart 1 and physical decay corrections required under subpart 7 must be performed by the authorized medical physicist.
A licensee must retain a record of each calibration according to part 4731.4518.
Minn. R. agency 144, ch. 4731, MEDICAL USE OF RADIOACTIVE MATERIAL, pt. 4731.4470
Statutory Authority: MS s 144.1202; 144.1203