THIS REPORT PREPARED PURSUANT TO DUTY IMPOSED BY A.A.C. R13-10-104(A)
ARIZONA DEPARTMENT OF PUBLIC SAFETY
STANDARD QUALITY ASSURANCE PROCEDURES ALCO SENSOR RBT AZ
STANDARD CALIBRATION CHECK PROCEDURE
AGENCY _______________________________________DATE ____________________
QA SPECIALIST _______________________________LOCATION __________________
RBT AZ SERIAL # _______________________ALCO SENSOR AZ SERIAL # __________
( ) 1. Have a standard alcohol concentration solution ready.
This may be a simulator (at 34° C ± 0.2° C) or a dry gas alcohol standard. Standard value: 0._____________AC.
( ) 2. Depress RBT AZ ON button.
Depress Time button.
Enter PIN #.
Depress zero button.
( ) 3. Follow RBT AZ and AS AZ display instructions.
( ) 4. Device temperature registers between 10° C and 40° C.
( ) 5. When AS AZ display reads "CHEK", introduce standard for 7 seconds; depress the MANUAL button on the
AS AZ at 5 seconds (while continuing to introduce the standard for another 2 seconds.)
( ) 6. Test results 0.___________ AC.
( ) 7. Remove test record when printout is complete.
( ) 8. Turn off RBT AZ.
COMMENTS _____________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
SIGNATURE ____________________________________________________________________
DPS Form Exh H-2 (Rev 05-01)
Ariz. Admin. Code tit. 13, ch. 10, art. 1, exh. H-2