Ariz. Admin. Code § tit. 13, ch. 10, art. 1, exh. H-2

Current through Register Vol. 30, No. 50, December 13, 2024
Exhibit H-2 - Standard Calibration Check Procedure Alco Sensor RBT AZ

THIS REPORT PREPARED PURSUANT TO DUTY IMPOSED BY A.A.C. R13-10-104(A)

ARIZONA DEPARTMENT OF PUBLIC SAFETY

STANDARD QUALITY ASSURANCE PROCEDURES ALCO SENSOR RBT AZ

STANDARD CALIBRATION CHECK PROCEDURE

AGENCY _______________________________________DATE ____________________

QA SPECIALIST _______________________________LOCATION __________________

RBT AZ SERIAL # _______________________ALCO SENSOR AZ SERIAL # __________

() 1. Have a standard alcohol concentration solution ready.

This may be a simulator (at 34° C ± 0.2° C) or a dry gas alcohol standard. Standard value: 0._____________AC.

() 2. Depress RBT AZ ON button.

Depress Time button.

Enter PIN #.

Depress zero button.

() 3. Follow RBT AZ and AS AZ display instructions.

() 4. Device temperature registers between 10° C and 40° C.

() 5. When AS AZ display reads "CHEK", introduce standard for 7 seconds; depress the MANUAL button on the

AS AZ at 5 seconds (while continuing to introduce the standard for another 2 seconds.)

() 6. Test results 0.___________ AC.

() 7. Remove test record when printout is complete.

() 8. Turn off RBT AZ.

COMMENTS _____________________________________________________________________

_______________________________________________________________________________

_______________________________________________________________________________

_______________________________________________________________________________

SIGNATURE ____________________________________________________________________

DPS Form Exh H-2 (Rev 05-01)

Ariz. Admin. Code tit. 13, ch. 10, art. 1, exh. H-2

New Exhibit H-2 made by final rulemaking at 12 A.A.R. 1916, effective 9:00 a.m., May 18, 2006 (Supp. 06-2).