Ariz. Admin. Code § tit. 13, ch. 10, art. 1, exh. H-1

Current through Register Vol. 30, No. 24, June 14, 2024
Exhibit H-1 - Standard Operational Procedure Alco Sensor RBT AZ

OPERATIONAL CHECKLIST

ARIZONA DEPARTMENT OF PUBLIC SAFETY

STANDARD OPERATIONAL PROCEDURE ALCO SENSOR RBT AZ

DUPLICATE BREATH TEST

SUBJECT NAME __________________________________DATE __________________________

AGENCY ___________________________________OPERATOR ___________________________

LOCATION _______________________________________________________________________

RBT AZ SERIAL # ____________________________ALCO SENSOR AZ SERIAL # ______________

TEST RESULTS 0. _________AC TIME _________

0. _________AC TIME _________

0. _________AC TIME _________

Immediately preceding administration of the tests, subject underwent at least a 15-minute deprivation period:

From

_______

to

_______

by

_______

(Time)

(Time)

(Name)

( ) 1. Depress RBT AZ ON button.

( ) 2. Depress zero set button, select subject or quick test.

( ) 3. Follow RBT AZ and AS AZ display instructions.

( ) 4. Enter case # &/or DL # if required.

( ) 5. Device temperature registers between 10° C and 40° C.

( ) 6.

a. If quick test, go to step 7.
b. If subject test, repeat steps 3 - 6 for duplicate test.
c. If the second subject test is not within 0.020 of the first test, repeat steps 3-6.
d. If the second subject test is within 0.020 of the first test, go to step 7.
e. If the third subject test, go to step 7.

( ) 7. Remove test record when printout is complete.

( ) 8. Turn off RBT AZ.

Note: Duplicate breath tests shall be administered at intervals of not less than 5 nor more than 10 minutes and the two consecutive tests shall agree within 0.020 alcohol concentration.

DPS Form Exh H-1 (Rev 05-01)

Ariz. Admin. Code tit. 13, ch. 10, art. 1, exh. H-1

New Exhibit H-1 made by final rulemaking at 12 A.A.R. 1916, effective 9:00 a.m., May 18, 2006 (Supp. 06-2).