Mo. Rev. Stat. § 190.092

Current with changes from the 2024 Legislative Session
Section 190.092 - Defibrillators, requirements - good faith immunity from civil liability, when - all basic life support ambulances and stretcher vans to be equipped with
1. This section shall be known and may be cited as the "Public Access to Automated External Defibrillator Act".
2. A person or entity that acquires an automated external defibrillator shall:
(1) Comply with all regulations governing the placement of an automated external defibrillator;
(2) Ensure that the automated external defibrillator is maintained and tested according to the operation and maintenance guidelines set forth by the manufacturer;
(3) Ensure that the automated external defibrillator is tested at least every two years and after each use; and
(4) Ensure that an inspection is made of all automated external defibrillators on the premises at least every ninety days for potential issues related to the operation of the device, including a blinking light or other obvious defect that may suggest tampering or that another problem has arisen with the functionality of the automated external defibrillator.
3. Any person who gratuitously and in good faith renders emergency care by use of or provision of an automated external defibrillator shall not be held liable for any civil damages or subject to any criminal penalty as a result of such care or treatment, unless the person acts in a willful and wanton or reckless manner in providing the care, advice, or assistance. The person who or entity that provides training to the person using an automated external defibrillator, the person or entity responsible for the site where the automated external defibrillator is located, and the person or entity that owns the automated external defibrillator shall likewise not be held liable for civil damages or subject to any criminal penalty resulting from the use of an automated external defibrillator.
4. All basic life support ambulances and stretcher vans operated in the state of Missouri shall be equipped with an automated external defibrillator and be staffed by at least one individual trained in the use of an automated external defibrillator.
5. The provisions of this section shall apply in all counties within the state and any city not within a county.

§ 190.092, RSMo

Amended by 2020 Mo. Laws, HB 1682,s A, eff. 7/13/2020.
L. 1998 H.B. 1668 § 190.375, A.L. 2002 S.B. 1107, A.L. 2004H.B. 1195 , A.L. 2009H.B. 103 , A.L. 2010H.B. 1977.