Table DHS 157.75
Half-Value Layer Requirements
X-Ray Tube Voltage in kilovolt peak | Half-Value Layer in mm Aluminum | |||
Design Operating Range | Mesured Operating Potential | Specified Dental Systems1 | I-Other X-Ray Systems2 | II-Other Diagnostic X-Ray Systems 3 |
Below 51 | 30 40 50 | 1.5 1.5 1.5 | 0.3 0.4 0.5 | 0.3 0.4 0.5 |
51 to 70 | 51 60 70 | 1.5 1.5 1.5 | 1.2 1.3 1.5 | 1.3 1.5 1.8 |
Above 70 | 71 80 90 100 110 120 130 140 150 | 2.1 2.3 2.5 2.7 3.0 3.2 3.5 3.8 4.1 | 2.1 2.3 2.5 2.7 3.0 3.2 3.5 3.8 4.1 | 2.5 2.9 3.2 3.6 3.9 4.3 4.7 5.0 5.4 |
1 Dental x-ray systems designed for use with intraoral image receptors and manufactured after December 1, 1980.
2 Dental x-ray systems designed for use with intraoral image receptors and manufactured before or on December 1, 1980, and all other x-ray systems subject to 21 CFR 1020 and manufactured before June 10, 2006.
3 All x-ray systems, except dental x-ray systems designed for use with intraoral image receptors, subject to 21 CFR 1020 and manufactured on or after June 10, 2006.
Wis. Admin. Code Department of Health Services DHS 157.75