Wash. Admin. Code § 246-240-614

Current through Register Vol. 25-01, January 1, 2025
Section 246-240-614 - Records of teletherapy, remote afterloader, and gamma stereotactic radiosurgery full calibrations
(1) A licensee shall maintain a record of the teletherapy unit, remote afterloader unit, and gamma stereotactic radiosurgery unit full calibrations required by WAC 246-240-369, 246-240-372, and 246-240-375 for three years.
(2) The record must include:
(a) The date of the calibration;
(b) The manufacturer's name, model number, and serial number of the teletherapy, remote afterloader, and gamma stereotactic radiosurgery units, the sources, and the instruments used to calibrate the units;
(c) The results and an assessment of the full calibrations;
(d) The results of the autoradiograph required for low dose-rate remote afterloader units; and
(e) The signature of the authorized medical physicist who performed the full calibration.

Wash. Admin. Code § 246-240-614

Amended by WSR 22-19-084, Filed 9/20/2022, effective 10/21/2022

Statutory Authority: RCW 70.98.050. 06-05-019, § 246-240-614, filed 2/6/06, effective 3/9/06.