Current through Register Vol. 25-01, January 1, 2025
Section 246-240-104 - Calibration of survey instruments(1) A licensee shall calibrate the survey instruments used to show compliance with this section and WAC 246-240-587 before first use, annually, and following a repair that affects the calibration. A licensee shall: (a) Calibrate all scales with readings up to 10 mSv (1000 mrem) per hour with a radiation source;(b) Calibrate two separated readings on each scale or decade that will be used to show compliance; and(c) Conspicuously note on the instrument the date of calibration.(2) A licensee may not use survey instruments if the difference between the indicated exposure rate and the calculated exposure rate is more than 20 percent.(3) A licensee shall retain a record of each survey instrument calibration in accordance with WAC 246-240-566.Wash. Admin. Code § 246-240-104
Amended by WSR 22-19-084, Filed 9/20/2022, effective 10/21/2022Statutory Authority: RCW 70.98.050. 06-05-019, § 246-240-104, filed 2/6/06, effective 3/9/06.