Wash. Admin. Code § 246-226-050

Current through Register Vol. 25-01, January 1, 2025
Section 246-226-050 - Dose limits

The CTDIvol for the following CT procedure on phantoms may not exceed the dose limits established in Table 1.

Table 1: Dose limits

CT ProcedurePhantom SizeDose Limit: CTDIvol (mGy)
Adult head 16 cm 80
Adult abdomen 32 cm 30
Pediatric head (one year old) 16 cm 40
Pediatric abdomen (40 pounds) 16 cm 20

Wash. Admin. Code § 246-226-050

Adopted by WSR 16-23-030, Filed 11/8/2016, effective 1/1/2017