Compliance with this section is required of healing arts registrants and is designed to ensure the patient and operator exposure is minimized, and to produce optimum image quality and diagnostic information.
THERMOMETER READINGS (DEGREES) | MINIMUM DEVELOPING TIMES (MINUTES) | ||
C | F | ||
27- | 80 | 2 | |
79 | 2 | ||
78 | 2 1/2 | ||
77 | 2 1/2 | ||
24- | 76 | 3 | |
75 | 3 | ||
74 | 3 1/2 | ||
73 | 3 1/2 | ||
22- | 72 | 4 | |
71 | 4 | ||
70 | 4 1/2 | ||
69 | 4 1/2 | ||
20- | 68 | 5 | |
67 | 5 1/2 | ||
66 | 5 1/2 | ||
65 | 6 | ||
18- | 64 | 6 1/2 | |
63 | 7 | ||
62 | 8 | ||
61 | 8 1/2 | ||
16- | 60 | 9 1/2 |
MINIMAL REQUIRED DEVELOPER TEMERATURE | PROCESSOR DEVELOPER IMMERSION TIME* | ||
°C | °F | Seconds | |
35 | 95 | 20 | |
34.5 | 94 | 21 | |
34 | 93 | 22 | |
33.5 | 92 | 23 | |
33 | 91 | 24 | |
32 | 90 | 25 | |
31.5 | 89 | 26 | |
31 | 88 | 27 | |
30.5 | 87 | 28 | |
30 | 86 | 29 | |
29.5 | 85 | 30 |
* | Immersion time only, no cross-over time included. |
The specified developer temperature and immersion time shall be posted in the dark room or on the automatic processor; and
Wash. Admin. Code § 246-225-150
Statutory Authority: RCW 70.98.050. 94-01-073, § 246-225-150, filed 12/9/93, effective 1/9/94. Statutory Authority: RCW 70.98.050 and 70.98.080. 91-15-083 (Order 183), § 246-225-150, filed 7/23/91, effective 8/23/91. Statutory Authority: RCW 43.70.040. 91-02-049 (Order 121), recodified as § 246-225-150, filed 12/27/90, effective 1/31/91. Statutory Authority: RCW 70.98.080. 83-19-050 (Order 2026), § 402-28-990, filed 9/16/83; Order 1084, Apendix C (codified as WAC 402-28-990), filed 1/14/76.