In addition to other requirements of this part, all diagnostic x-ray systems shall meet the following requirements:
"WARNING: This x-ray unit may be dangerous to patient and operator unless safe exposure factors, operating instructions, and maintenance schedules are observed."
TABLE 1 | ||||
X-Ray Tube Voltage (kilovolt peak) | ||||
Design Operating Range | Measured Operating Potential | Minimum HVL (mm in Aluminum) | ||
Specified Dental Systems1 | I-Other X-Ray Systems2 | II-Other X-Ray Systems3 | ||
Below 51 | 30 | 1.5 | 0.3 | 0.3 |
40 | 1.5 | 0.4 | 0.4 | |
50 | 1.5 | 0.5 | 0.5 | |
51 to 70 | 51 | 1.5 | 1.2 | 1.3 |
60 | 1.5 | 1.3 | 1.5 | |
70 | 1.5 | 1.5 | 1.8 | |
Above 70 | 71 | 2.1 | 2.1 | 2.5 |
80 | 2.3 | 2.3 | 2.9 | |
90 | 2.5 | 2.5 | 3.2 | |
100 | 2.7 | 2.7 | 3.6 | |
110 | 3.0 | 3.0 | 3.9 | |
120 | 3.2 | 3.2 | 4.3 | |
130 | 3.5 | 3.5 | 4.7 | |
140 | 3.8 | 3.8 | 5.0 | |
150 | 4.1 | 4.1 | 5.4 | |
1 Dental x-ray systems designed for use with intraoral image receptors and manufactured after December 1, 1980. 2 Dental x-ray systems designed for use with intraoral image receptors and manufactured before or on December 1, 1980, and all other x-ray systems subject to this section and manufactured before June 10, 2006. 3 All x-ray systems, except dental x-ray systems designed for use with intraoral image receptors, subject to this section and manufactured on or after June 10, 2006. |
TABLE 2 | |
Item | Maximum Aluminum Equivalent (mm) |
Front panel(s) of cassette holders (total of all) | 1.2 |
Film panel(s) of film changer (total of all) | 1.2 |
Cradle | 2.3 |
Tabletop, stationary, without articulated joints | 1.2 |
Tabletop, movable, without articulated joints (including stationary subtop) | 1.7 |
Tabletop, with radiolucent panel having one articulated joint | 1.7 |
Tabletop, with radiolucent panel having two or more articulated joints | 2.3 |
Tabletop, cantilevered | 2.3 |
Tabletop, radiation therapy simulator | 5.0 |
12 Va. Admin. Code § 5-481-1601
Statutory Authority: § 32.1-229 of the Code of Virginia.