Current through Bulletin No. 2024-21, November 1, 2024
Section R315-261-1054 - Air Emission Standards: Pressure Relief Devices in Gas/Vapor Service(a) Except during pressure releases, each pressure relief device in gas/vapor service shall be operated with no detectable emissions, as indicated by an instrument reading of less than 500 ppm above background, as measured by the method specified in Subsection R315-261-1063(c).(b)(1) After each pressure release, the pressure relief device shall be returned to a condition of no detectable emissions, as indicated by an instrument reading of less than 500 ppm above background, as soon as practicable, but no later than 5 calendar days after each pressure release, except as provided in Section R315-261-1059.(2) No later than 5 calendar days after the pressure release, the pressure relief device shall be monitored to confirm the condition of no detectable emissions, as indicated by an instrument reading of less than 500 ppm above background, as measured by the method specified in Subsection R315-261-1063(c).(c) Any pressure relief device that is equipped with a closed-vent system capable of capturing and transporting leakage from the pressure relief device to a control device as described in Section R315-261-1060 is exempt from the requirements of Subsection R315-261-1054(a) and (b).Utah Admin. Code R315-261-1054
Adopted by Utah State Bulletin Number 2016-9, effective 4/15/2016