In addition to other requirements of R313-28, all diagnostic x-ray systems shall meet the following requirements:
TABLE I | |||
DESIGN OPERATING RANGE (KILO VOLTS PEAK | MEASURED POTENTIAL (KILOVOLTS PEAK) | DENTAL INTRA-ORAL MANUFACTURED BEFORE AUGUST 1, 1974 AND ON OR AFTER DECEMBER 1, 1980 | ALL OTHER DIAGNOSTIC X-RAY SYSTEMS |
Below 51 | 30 | (use prohibited) | 0.3 |
40 | (use prohibited) | 0.4 | |
50 | 1.5 | 0.5 | |
51 | 1.5 | 1.2 | |
60 | 1.5 | 1.3 | |
70 | 1.5 | 1.5 | |
Above 70 | 71 | 2.1 | 2.1 |
80 | 2.3 | 2.3 | |
90 | 2.5 | 2.5 | |
100 | 2.7 | 2.7 | |
110 | 3.0 | 3.0 | |
120 | 3.2 | 3.2 | |
130 | 3.5 | 3.5 | |
140 | 3.8 | 3.8 | |
150 | 4.1 | 4.1 |
Utah Admin. Code R313-28-35