Current through October 22, 2024
Section 0400-20-09-.20 - TESTS AND SURVEYSEach registrant shall:
(1) Test all safety and warning devices including interlocks at intervals not to exceed 3 months to determine that they are functioning properly;(2) In conjunction with initial operation and after changes have been made in shielding, operating parameters, equipment, or occupancy of adjacent areas, make a survey as required in Rule 0400-20-05-.70;(3) Test any interlock which has been bypassed or otherwise prevented from operation, when such interlock is returned to use, to determine if it is functioning properly; and(4) Have a survey made of each radiation area upon the initial entry by personnel into these areas following the operation of accelerator. The registrant shall not be required to make a record of the survey required by this paragraph.Tenn. Comp. R. & Regs. 0400-20-09-.20
Original rule filed February 22, 2012; effective May 22, 2012.Authority: T.C.A. §§ 68-202-101 et seq., 68-202-201 et seq. and 4-5-201 et seq.