Section 0400-20-09-.20 - TESTS AND SURVEYS

Each registrant shall:

(1) Test all safety and warning devices including interlocks at intervals not to exceed 3 months to determine that they are functioning properly;
(2) In conjunction with initial operation and after changes have been made in shielding, operating parameters, equipment, or occupancy of adjacent areas, make a survey as required in Rule 0400-20-05-.70;
(3) Test any interlock which has been bypassed or otherwise prevented from operation, when such interlock is returned to use, to determine if it is functioning properly; and
(4) Have a survey made of each radiation area upon the initial entry by personnel into these areas following the operation of accelerator. The registrant shall not be required to make a record of the survey required by this paragraph.