Section 0400-20-07-.58 - THERAPY-RELATED COMPUTER SYSTEMS(1) The licensee shall perform or shall verify and maintain documentation of acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of:(a) The source-specific input parameters required by the dose calculation algorithm;(b) The accuracy of dose, dwell time, and treatment time calculations at representative points;(c) The accuracy of isodose plots and graphic displays; and(d) The accuracy of the software used to determine sealed source positions from radiographic images.Tenn. Comp. R. & Regs. 0400-20-07-.58