Tenn. Comp. R. & Regs. 0400-20-07-.108

Current through October 22, 2024
Section 0400-20-07-.108 - RECORDS OF ADDITIONAL TECHNICAL REQUIREMENTS FOR MOBILE REMOTE AFTERLOADER UNITS
(1) A licensee shall retain a record of each check for mobile remote afterloader units required by Rule 0400-20-07-.75 for 3 years.
(2) The record must include:
(a) The date of the check;
(b) The manufacturer's name, model number, and serial number of the remote afterloader unit;
(c) Notations accounting for all sources before the licensee departs from a facility;
(d) Notations indicating the operability of each entrance door electrical interlock, radiation monitors, source exposure indicator lights, viewing and intercom system, applicators, source transfer tubes, and transfer tube applicator interfaces, and source positioning accuracy; and
(e) The signature of the individual who performed the check.

Tenn. Comp. R. & Regs. 0400-20-07-.108

Original rule filed February 22, 2012; effective May 22, 2012.

Authority: T.C.A. §§ 68-202-101 et seq., 68-202-201 et seq. and 4-5-201 et seq.