Current through October 22, 2024
Section 0400-20-07-.106 - RECORDS OF PERIODIC SPOT-CHECKS FOR REMOTE AFTERLOADER UNITS(1) A licensee shall retain a record of each spot-check for remote afterloader units required by Rule 0400-20-07-.73 for 3 years.(2) The record must include, as applicable: (a) The date of the spot-check; .(b) The manufacturer's name, model number, and serial number for the remote afterloader unit and source;(c) An assessment of timer accuracy;(d) Notations indicating the operability of each entrance door electrical interlock, radiation monitors, source exposure indicator lights, viewing and intercom systems, and clock and decayed source activity in the unit's computer; and(e) The name of the individual who performed the periodic spot-check and the signature of the authorized medical physicist who reviewed the record of the spot-check.Tenn. Comp. R. & Regs. 0400-20-07-.106
Original rule filed February 22, 2012; effective May 22, 2012.Authority: T.C.A. §§ 68-202-101 et seq., 68-202-201 et seq. and 4-5-201 et seq.