Current through Register Vol. 48, No. 10, October 25, 2024
Section 61-64.VIII.RHB 8.10 - Inspections and MaintenanceEach registrant shall ensure that checks for obvious defects in radiation machines are made at the beginning of each day of equipment use.
8.10.1 At least annually, each registrant shall inspect and repair components associated with radiation safety of the machines. Records of inspection and maintenance shall be maintained for the Department's inspection.8.10.2 If any inspection conducted by the registrant reveals damage to the components affecting radiation safety, the radiation machine shall not be used and shall be labeled as defective until repaired.S.C. Code Regs. § 61-64.VIII.RHB 8.10
Replaced and amended by State Register Volume 40, Issue No. 06, eff. 6/24/2016; State Register Volume 47, Issue No. 05, eff. 5/26/2023.