Current through Register Vol. 48, No. 10, October 25, 2024
Section 61-64.VII.RHB 7.6 - Additional Requirements for Enclosed Beam X-ray EquipmentTo include stationary, transportable, mobile, and portable units.
7.6.1 The radiation source, sample, detector, and analyzing crystal (if used) shall be enclosed in a chamber or coupled chambers that cannot be entered by any part of the body during normal operation.7.6.2 The sample chamber closure shall be interlocked with the x-ray tube high voltage supply or a shutter in the primary beam so that no x-ray beam can enter the sample chamber while it is open unless the interlock has been conspicuously and deliberately defeated. The interlock required by this section shall be of fail-safe design or adequate administrative controls shall be exercised to ensure operations will not continue without a properly functioning interlock.S.C. Code Regs. § 61-64.VII.RHB 7.6
Replaced and amended by State Register Volume 40, Issue No. 06, eff. 6/24/2016; State Register Volume 47, Issue No. 05, eff. 5/26/2023.