These tables are used to determine the mean glandular dose in milligrays delivered by 25.9 mC/kg (or millirad) delivered by one Roentgen (1 R) in air incident on a 4.2 centimeter thickness compressed breast of average density (fifty percent (50%) adipose and fifty percent (50%) glandular tissue). Linear extrapolation or interpolation shall be made for any HVL not listed. To convert from entrance exposure in air in roentgens to mean glandular breast dose in millirads, multiply the entrance exposure by the factor shown in the table for the appropriate kVp and beam quality (HVL) combination.
GLANDULAR DOSE (IN mrad) FOR 1 ROENTGEN ENTRANCE EXPOSURE TO A 4.2-CM BREAST THICKNESS---50% ADIPOSE 50% GLANDULAR BREAST TISSUE---USING A Mo/Mo TARGET-FILTER COMBINATION*
X-ray Tube Voltage (kVp) | W/Al | |||||||||||
Target-Filter | ||||||||||||
HVL | 23 | 24 | 25 | 26 | 27 | 28 | 29 | 30 | 31 | 32 | 33 | Combination |
0.23 | 116 | |||||||||||
0.24 | 121 | 124 | ||||||||||
0.25 | 126 | 129 | 131 | |||||||||
0.26 | 130 | 133 | 135 | 138 | ||||||||
0.27 | 135 | 138 | 140 | 142 | 143 | |||||||
0.28 | 140 | 142 | 144 | 146 | 147 | 149 | ||||||
0.29 | 144 | 146 | 148 | 150 | 151 | 153 | 154 | |||||
0.30 | 149 | 151 | 153 | 155 | 156 | 157 | 158 | 159 | 170 | |||
0.31 | 154 | 156 | 157 | 159 | 160 | 161 | 162 | 163 | 164 | 175 | ||
0.32 | 158 | 160 | 162 | 163 | 164 | 166 | 167 | 168 | 168 | 170 | 171 | 180 |
0.33 | 163 | 165 | 166 | 168 | 169 | 170 | 171 | 173 | 173 | 174 | 175 | 185 |
0.34 | 168 | 170 | 171 | 172 | 173 | 174 | 175 | 176 | 177 | 178 | 179 | 190 |
0.35 | 174 | 175 | 176 | 177 | 178 | 179 | 180 | 181 | 182 | 183 | 194 | |
0.36 | 179 | 181 | 182 | 183 | 184 | 185 | 185 | 186 | 187 | 199 | ||
0.37 | 185 | 186 | 187 | 188 | 189 | 190 | 191 | 191 | 204 | |||
0.38 | 190 | 191 | 192 | 193 | 194 | 195 | 195 | 208 | ||||
0.39 | 196 | 197 | 198 | 198 | 199 | 200 | 213 | |||||
0.40 | 201 | 202 | 203 | 204 | 204 | 217 | ||||||
0.41 | 206 | 207 | 208 | 208 | 221 | |||||||
0.42 | 211 | 212 | 212 | 225 | ||||||||
0.43 | 215 | 216 | 230 | |||||||||
0.44 | 220 | 234 | ||||||||||
0.45 | 238 |
GLANDULAR DOSE (IN mrad) FOR 1 ROENTGEN ENTRANCE EXPOSURE TO A 4.2-CM BREAST THICKNESS ---50% ADIPOSE 50% GLANDULAR BREAST TISSUE ---USING A Mo/Rh TARGET-FILTER COMBINATION*
X-ray Tube Voltage (kVp) | |||||||||||
HVL | 25 | 26 | 27 | 28 | 29 | 30 | 31 | 32 | 33 | 34 | 35 |
0.28 | 149 | 151 | 154 | ||||||||
0.29 | 154 | 156 | 158 | 159 | |||||||
0.30 | 158 | 160 | 162 | 162 | 163 | ||||||
0.31 | 163 | 164 | 166 | 166 | 166 | 167 | 167 | ||||
0.32 | 167 | 169 | 171 | 171 | 171 | 171 | 172 | 172 | |||
0.33 | 171 | 173 | 175 | 176 | 176 | 176 | 176 | 177 | |||
0.34 | 176 | 178 | 179 | 179 | 180 | 180 | 180 | 181 | 181 | ||
0.35 | 180 | 181 | 183 | 183 | 184 | 185 | 185 | 186 | 187 | ||
0.36 | 185 | 186 | 187 | 187 | 188 | 188 | 189 | 190 | 191 | 191 | |
0.37 | 189 | 190 | 191 | 191 | 192 | 193 | 193 | 194 | 195 | 195 | |
0.38 | 193 | 194 | 196 | 196 | 197 | 197 | 197 | 198 | 199 | 199 | 200 |
0.39 | 198 | 199 | 200 | 200 | 201 | 201 | 202 | 202 | 203 | 203 | 204 |
0.40 | 202 | 203 | 204 | 204 | 205 | 205 | 206 | 207 | 208 | 208 | 208 |
0.41 | 206 | 207 | 208 | 208 | 209 | 209 | 210 | 211 | 212 | 212 | 212 |
0.42 | 211 | 211 | 212 | 212 | 213 | 213 | 214 | 215 | 216 | 216 | 217 |
0.43 | 215 | 216 | 217 | 217 | 218 | 218 | 219 | 219 | 220 | 220 | 221 |
0.44 | 220 | 220 | 221 | 221 | 222 | 222 | 223 | 223 | 224 | 224 | 225 |
0.45 | 224 | 224 | 225 | 225 | 226 | 226 | 227 | 227 | 228 | 228 | 229 |
0.46 | 228 | 229 | 229 | 230 | 231 | 231 | 232 | 233 | 233 | 234 | |
0.47 | 233 | 233 | 234 | 235 | 235 | 236 | 237 | 237 | 238 | ||
0.48 | 238 | 238 | 239 | 240 | 240 | 241 | 241 | 242 | 242 | ||
0.49 | 242 | 243 | 243 | 244 | 244 | 245 | 245 | 246 | |||
0.50 | 247 | 247 | 248 | 248 | 249 | 250 | 251 | ||||
0.51 | 251 | 252 | 253 | 254 | 254 | 255 | |||||
0.52 | 257 | 257 | 258 | 258 | 259 | ||||||
0.53 | 261 | 261 | 262 | 263 | 264 | ||||||
0.54 | 265 | 266 | 267 | 268 | |||||||
0.55 | 269 | 270 | 271 | 272 | |||||||
0.56 | 275 | 276 | 276 | ||||||||
0.57 | 279 | 280 | 281 | ||||||||
0.58 | 284 | 285 | |||||||||
0.59 | 288 | 289 | |||||||||
0.60 | 293 |
GLANDULAR DOSE (IN mrad) FOR 1 ROENTGEN ENTRANCE EXPOSURE TO A 4.2-CM BREAST THICKNESS ---50% ADIPOSE 50% GLANDULAR BREAST TISSUE ---USING A Rh/Rh TARGET-FILTER COMBINATION*
X-ray Tube Voltage (kVp) | |||||||||||
HVL | 25 | 26 | 27 | 28 | 29 | 30 | 31 | 32 | 33 | 34 | 35 |
0.28 | 150 | 155 | 159 | ||||||||
0.29 | 155 | 160 | 164 | 168 | |||||||
0.30 | 160 | 164 | 168 | 172 | 176 | ||||||
0.31 | 165 | 168 | 172 | 174 | 180 | 182 | |||||
0.32 | 169 | 173 | 177 | 181 | 184 | 186 | 188 | ||||
0.33 | 174 | 178 | 181 | 185 | 188 | 190 | 192 | ||||
0.34 | 179 | 183 | 186 | 190 | 193 | 195 | 196 | 199 | |||
0.35 | 184 | 187 | 190 | 194 | 197 | 199 | 201 | 203 | |||
0.36 | 189 | 192 | 195 | 198 | 201 | 204 | 205 | 207 | 209 | ||
0.37 | 193 | 196 | 199 | 202 | 205 | 207 | 209 | 211 | 213 | ||
0.38 | 198 | 201 | 204 | 207 | 209 | 211 | 213 | 215 | 217 | 219 | 221 |
0.39 | 203 | 206 | 208 | 211 | 214 | 216 | 217 | 219 | 221 | 223 | 224 |
0.40 | 208 | 211 | 213 | 216 | 218 | 220 | 221 | 223 | 224 | 226 | 228 |
0.41 | 213 | 215 | 217 | 220 | 222 | 224 | 225 | 227 | 228 | 230 | 232 |
0.42 | 218 | 220 | 222 | 224 | 226 | 228 | 229 | 231 | 232 | 234 | 236 |
0.43 | 222 | 224 | 226 | 228 | 230 | 232 | 233 | 235 | 236 | 238 | 240 |
0.44 | 227 | 229 | 231 | 233 | 235 | 237 | 238 | 239 | 240 | 242 | 243 |
0.45 | 232 | 234 | 235 | 237 | 239 | 241 | 242 | 243 | 244 | 246 | 247 |
0.46 | 239 | 241 | 243 | 245 | 246 | 247 | 248 | 250 | 251 | ||
0.47 | 247 | 249 | 250 | 251 | 252 | 254 | 255 | ||||
0.48 | 251 | 253 | 254 | 255 | 256 | 258 | 259 | ||||
0.49 | 257 | 258 | 259 | 260 | 261 | 262 | |||||
0.50 | 261 | 262 | 263 | 264 | 265 | 266 | |||||
0.51 | 266 | 267 | 268 | 269 | 270 | ||||||
0.52 | 270 | 271 | 272 | 273 | 274 | ||||||
0.53 | 275 | 276 | 276 | 277 | 278 | ||||||
0.54 | 279 | 280 | 280 | 281 | |||||||
0.55 | 283 | 284 | 284 | 285 | |||||||
0.56 | 288 | 288 | 289 | ||||||||
0.57 | 292 | 293 | |||||||||
0.58 | 296 | 297 | |||||||||
0.59 | 300 | ||||||||||
0.60 | 304 |