S.C. Code Regs. § § 61-64.IX.RHB 9.4

Current through Register Vol. 48, 12, December 27, 2024
Section 61-64.IX.RHB 9.4 - Shielding

Prior to installation or replacement, the registrant shall submit a floor plan and equipment arrangement which has been prepared by a registered Class II vendor and submitted to the Department for review and acceptance.

9.4.1 The floor plan must include, at a minimum:
9.4.1.1 The proposed location of the system;
9.4.1.2 Surrounding and adjacent areas with occupancies;
9.4.1.3 General direction of the useful beam; and
9.4.1.4 Location of the control panel and operator.
9.4.2 An inspection zone shall be established around the personnel security screening system where bystanders are prohibited during the operation of the device. A means shall be provided for any operator responsible for initiating a scan to maintain full visual surveillance of the inspection zone. The registrant shall ensure only the scanned individual is within two meters (2 m) of the scanner when in operation.
9.4.3 The Department may require a shielding plan, as described in RHB 4.4.

S.C. Code Regs. § 61-64.IX.RHB 9.4

Added by State Register Volume 47, Issue No. 05, eff. 5/26/2023.