S.C. Code Regs. § § 61-63.IV.L.RHA 4.112

Current through Register Vol. 48, 12, December 27, 2024
Section 61-63.IV.L.RHA 4.112 - Records of Periodic Spot-checks for Remote Afterloader Units
4.112.1 A licensee shall retain a record of each spot-check for remote afterloader units required by RHA 4.68 for 3 years.
4.112.2 The record must include, as applicable--
4.112.2.1 The date of the spot-check;
4.112.2.2 The manufacturer's name, model number, and serial number for the remote afterloader unit and source;
4.112.2.3 An assessment of timer accuracy;
4.112.2.4 Notations indicating the operability of each entrance door electrical interlock, radiation monitors, source exposure indicator lights, viewing and intercom systems, and clock and decayed source activity in the unit's computer; and
4.112.2.5 The name of the individual who performed the periodic spot-check and the signature of the authorized medical physicist who reviewed the record of the spot-check.
4.112.3 A licensee shall retain a copy of the procedures required by RHA 4.68.2 until the licensee no longer possesses the remote afterloader unit.

S.C. Code Regs. § 61-63.IV.L.RHA 4.112