Current through Register Vol. 48, 12, December 27, 2024
Section 61-63.IV.L.RHA 4.110 - Records of Teletherapy, Remote Afterloader, and Gamma Stereotactic Radiosurgery Full Calibrations4.110.1 A licensee shall maintain a record of the teletherapy unit, remote afterloader unit, and gamma stereotactic radiosurgery unit full calibrations required by RHA 4.64, 4.65 and 4.66 for 3 years.4.110.2 The record must include-- 4.110.2.1 The date of the calibration;4.110.2.2 The manufacturer's name, model number, and serial number of the teletherapy, remote afterloader, and gamma stereotactic radiosurgery unit(s), the source(s), and the instruments used to calibrate the unit(s);4.110.2.3 The results and an assessment of the full calibrations;4.110.2.4 The results of the autoradiograph required for low dose-rate remote afterloader units; and4.110.2.5 The signature of the authorized medical physicist who performed the full calibration.S.C. Code Regs. § 61-63.IV.L.RHA 4.110