Section 61-63.IV.H.RHA 4.68 - Periodic Spot-Checks for Remote Afterloader Units4.68.1 A licensee authorized to use a remote afterloader unit for medical use shall perform spot-checks of each remote afterloader facility and on each unit-- 4.68.1.1 Before the first use of a high dose-rate, medium dose-rate, or pulsed dose-rate remote afterloader unit on a given day;4.68.1.2 Before each patient treatment with a low dose-rate remote afterloader unit; and4.68.1.3 After each source installation.4.68.2 A licensee shall perform the measurements required by RHA 4.68.1 in accordance with written procedures established by the authorized medical physicist. That individual need not actually perform the spot check measurements.4.68.3 A licensee shall have the authorized medical physicist review the results of each spot-check within 15 days. The authorized medical physicist shall notify the licensee as soon as possible in writing of the results of each spot-check.4.68.4 To satisfy the requirements of RHA 4.68.1, spot-checks must, at a minimum, assure proper operation of-- 4.68.4.1 Electrical interlocks at each remote afterloader unit room entrance;4.68.4.2 Source exposure indicator lights on the remote afterloader unit, on the control console, and in the facility;4.68.4.3 Viewing and intercom systems in each high dose-rate, medium dose-rate, and pulsed dose-rate remote afterloader facility;4.68.4.4 Emergency response equipment;4.68.4.5 Radiation monitors used to indicate the source position;4.68.4.7 Clock (date and time) in the unit's computer; and4.68.4.8 Decayed source(s) activity in the unit's computer.4.68.5 If the results of the checks required in RHA 4.68.4 indicate the malfunction of any system, a licensee shall lock the control console in the off position and not use the unit except as may be necessary to repair, replace, or check the malfunctioning system.4.68.6 A licensee shall retain a record of each check required by RHA 4.68.4 and a copy of the procedures required by RHA 4.68.2 in accordance with RHA 4.112.S.C. Code Regs. § 61-63.IV.H.RHA 4.68