Section 61-63.IV.F.RHA 4.53 - Therapy-Related Computer Systems4.53.1 The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of: 4.53.1.1 The source-specific input parameters required by the dose calculation algorithm;4.53.1.2 The accuracy of dose, dwell time, and treatment time calculations at representative points;4.53.1.3 The accuracy of isodose plots and graphic displays; and4.53.1.4 The accuracy of the software used to determine sealed source positions from radiographic images.S.C. Code Regs. § 61-63.IV.F.RHA 4.53