S.C. Code Regs. § § 61-63.IV.F.RHA 4.53

Current through Register Vol. 48, 12, December 27, 2024
Section 61-63.IV.F.RHA 4.53 - Therapy-Related Computer Systems
4.53.1 The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of:
4.53.1.1 The source-specific input parameters required by the dose calculation algorithm;
4.53.1.2 The accuracy of dose, dwell time, and treatment time calculations at representative points;
4.53.1.3 The accuracy of isodose plots and graphic displays; and
4.53.1.4 The accuracy of the software used to determine sealed source positions from radiographic images.

S.C. Code Regs. § 61-63.IV.F.RHA 4.53