Current through Register Vol. 48, 12, December 27, 2024
Section 61-58.5.G - Microbiological Contaminant Sampling and Analytical RequirementsThese sampling and analytical requirements shall apply to community and non-community water systems. Analytical methods used to comply with Section F above, shall be made using EPA-approved methods listed in 40 CFR 141.
(1) Routine Monitoring. (a) Community and non-community water systems shall collect total coliform samples at sites which are representative of water throughout the distribution system according to a written sample siting plan. These plans are subject to Department review and revision.(b) The monitoring frequency for total coliforms for community water systems is based on the population served by the system, as follows: Population Served | Minimum # of Samples Per Month |
25 to 1,000 [FN1] | 1 |
1,001 to 2,500 | 2 |
2,501 to 3,300 | 3 |
3,301 to 4,100 | 4 |
4,101 to 4,900 | 5 |
4,901 to 5,800 | 6 |
5,801 to 6,700 | 7 |
6,701 to 7,600 | 8 |
7,601 to 8,500 | 9 |
8,501 to 12,900 | 10 |
12,901 to 17,200 | 15 |
17,201 to 21,500 | 20 |
21,501 to 25,000 | 25 |
25,001 to 33,000 | 30 |
33,001 to 41,000 | 40 |
41,001 to 50,000 | 50 |
50,001 to 59,000 | 60 |
59,001 to 70,000 | 70 |
70,001 to 83,000 | 80 |
83,001 to 96,000 | 90 |
96,001 to 130,000 | 100 |
130,001 to 220,000 | 120 |
220,001 to 320,000 | 150 |
320,001 to 450,000 | 180 |
450,001 to 600,000 | 210 |
600,001 to 780,000 | 240 |
780,001 to 970,000 | 270 |
970,001 to 1,230,000 | 300 |
1,230,001 to 1,520,000 | 330 |
1,520,001 to 1,850,000 | 360 |
1,850,001 to 2,270,000 | 390 |
2,270,001 to 3,020,000 | 420 |
3,020,001 to 3,960,000 | 450 |
3,960,000 or more | 480 |
[FN1] Includes public water systems which have at least fifteen (15) service connections, but serve fewer than twenty-five (25) persons. |
If a community water system serving twenty-five (25) to one-thousand (1,000) persons has no history of total coliform contamination in its current configuration and a sanitary survey conducted in the past five years shows that the system is supplied solely by a protected groundwater source and is free of sanitary defects, the Department may reduce the monitoring frequency specified above, except that in no case may the Department reduce the monitoring frequency to less than one sample per quarter. The Department must approve the reduced monitoring frequency in writing.
(ii) Community water systems shall make at a minimum one fecal or total coliform density measurement each day from the raw water source, and one coliform density or presence/absence measurement from the finished water, if treating surface water. This requirement may be waived by the Department on a case-by-case basis if a public water supply can demonstrate that such monitoring is unnecessary.(c) The monitoring frequency for total coliforms for non-community water systems is as follows: (i) A non-community water system using only ground water (except ground water under the direct influence of surface water) and serving one-thousand (1,000) persons or fewer shall monitor each calendar quarter that the system provides water to the public, except that the Department may reduce this monitoring frequency, in writing, if a sanitary survey shows that the system is free of sanitary defects. Beginning June 29, 1994, the Department cannot reduce the monitoring frequency for a non-community water system using only ground water (except ground water under the direct influence of surface water) and serving one-thousand (1,000) persons or fewer to less than once per year.(ii) A non-community water system using only ground water (except ground water under the direct influence of surface water) and serving more than one-thousand (1,000) persons during any month shall monitor at the same frequency as a like-sized community water system, as specified in paragraph (1)(b) of this section, except that the Department may reduce this monitoring frequency, in writing, for any month the system serves one-thousand (1,000) persons or fewer. The Department cannot reduce the monitoring frequency to less than once per year. For systems using ground water under the direct influence of surface water, paragraph (1)(c)(iv) of this section applies.(iii) A non-community water system using surface water, in total or in part, shall monitor at the same frequency as a like-sized community water system, as specified in paragraph (1)(b) of this section, regardless of the number of persons it serves.(iv) A non-community water system using ground water under the direct influence of surface water shall monitor at the same frequency as a like-sized community water system, as specified in paragraph (1)(b) of this section. The system shall begin monitoring at this frequency beginning six (6) months after the Department determines that the ground water is under the direct influence of surface water.(d) The community or non-community water system shall collect samples at regular time intervals throughout the month, except that a system which uses ground water (except ground water under the direct influence of surface water), and serves 4,900 persons or fewer, may collect all required samples on a single day if they are taken from different sites.(e) A community or non-community water systems that uses surface water or ground water under the direct influence of surface water and does not practice filtration in compliance with R.61-58.10 shall collect at least one sample near the first service connection each day the turbidity level of the source water, measured as specified in R.61-58.10.F(2)(b), exceeds 1 NTU. This sample shall be analyzed for the presence of total coliforms. When one or more turbidity measurements in any day exceed 1 NTU, the system shall collect this coliform sample within 24 hours of the first exceedance, unless the Department determines that the system, for logistical reasons outside the system's control, cannot have the sample analyzed within thirty (30) hours of collection. Sample results from this coliform monitoring shall be included in determining compliance with the MCL for total coliforms in Section F above.(f) Special purpose samples, such as those taken to determine whether disinfection practices are sufficient following pipe placement, replacement, or repair, shall not be used to determine compliance with the MCL for total coliforms in Section F above. Repeat samples taken pursuant to paragraph (2) of this section are not considered special purpose samples, and shall be used to determine compliance with the MCL for total coliforms in Section F above.(2) Repeat Monitoring. (a) If a routine sample is total coliform-positive, the community or non-community water system shall collect a set of repeat samples within twenty-four (24) hours of being notified of the positive result. A system which collects more than one routine sample per month shall collect no fewer than three repeat samples for each total coliform-positive sample found. A system which collects one routine sample per month or fewer shall collect no fewer than four repeat samples for each total coliform-positive sample found. The Department may extend the twenty-four (24) hour limit on a case-by-case basis if the system has a logistical problem in collecting the repeat samples within twenty-four (24) hours that is beyond its control. In the case of an extension, the Department shall specify how much time the system has to collect the repeat samples.(b) The system shall collect at least one repeat sample from the sampling tap where the original total coliform-positive sample was taken, and at least one repeat sample at a tap within five service connections upstream and at least one repeat sample at a tap within five (5) service connections downstream of the original sampling site. If a total coliform-positive sample is at the end of the distribution system, or one away from the end of the distribution system, the Department may waive the requirement to collect at least one repeat sample upstream or downstream of the original sampling site.(c) The system shall collect all repeat samples on the same day, except that the Department may allow a system with a single service connection to collect the required set of repeat samples over a four-day period or to collect a larger volume repeat sample(s) in one or more sample containers of any size, as long as the total volume collected is at least 400 ml (300 ml for systems which collect more than one routine sample per month).(d) If one or more repeat samples in the set is total coliform-positive, the water system shall collect an additional set of repeat samples in the manner specified in paragraphs (2)(a) through (c) of this section. The additional samples shall be collected within twenty-four (24) hours of being notified of the positive result, unless the Department extends the limit as provided in paragraph (2)(a) of this section. The system shall repeat this process until either total coliforms are not detected in one complete set of repeat samples or the system determines that the MCL for total coliforms in Section F above, has been exceeded and notifies the Department.(e) If a system collecting fewer than five routine samples per month has one or more total coliform-positive samples and the Department does not invalidate the sample(s) under paragraph (3) of this section, it shall collect at least five routine samples during the next month the system provides water to the public, except that the Department may waive this requirement if the conditions of paragraph (2)(e)(i) or (ii) of this section are met. The Department cannot waive the requirement for a system to collect repeat samples in paragraphs (2)(a) through (d) of this section. (i) The Department may waive the requirement to collect five routine samples the next month the system provides water to the public if the Department, or an agent approved by the Department, performs a site visit before the end of the next month the system provides water to the public. Although a sanitary survey need not be performed, the site visit shall be sufficiently detailed to allow the Department to determine whether additional monitoring and/or any corrective action is needed. The Department cannot approve an employee of the system to perform this site visit, even if the employee is an agent approved by the Department to perform sanitary surveys.(ii) The Department may waive the requirement to collect five routine samples the next month the system provides water to the public if the Department has determined why the sample was total coliform-positive and establishes that the system has corrected the problem or will correct the problem before the end of the next month the system serves water to the public. In this case, the Department shall document this decision to waive the following month's additional monitoring requirement in writing, have it approved and signed by the supervisor of the Department official who recommends such a decision, and make this document available to the EPA and public. The written documentation shall describe the specific cause of the total coliform-positive sample and what action the system has taken and/or will take to correct this problem. The Department cannot waive the requirement to collect five routine samples the next month the system provides water to the public solely on the grounds that all repeat samples are total coliform-negative. Under this paragraph, a system shall still take at least one routine sample before the end of the next month it serves water to the public and use it to determine compliance with the MCL for total coliforms in R.61-58.5.F, unless the Department has determined that the system has corrected the contamination problem before the system took the set of repeat samples required in paragraphs (2)(a) through (d) of this section, and all repeat samples were total coliform-negative.(f) After a system collects a routine sample and before it learns the results of the analysis of that sample, if it collects another routine sample(s) from within five adjacent service connections of the initial sample, and the initial sample, after analysis, is found to contain total coliforms, then the system may count the subsequent sample(s) as a repeat sample instead of as a routine sample.(g) Results of all routine and repeat samples not invalidated by the Department shall be included in determining compliance with the MCL for total coliforms in Section F above.(3) Invalidation of total coliform samples. A total coliform-positive sample invalidated under this paragraph does not count towards meeting the minimum monitoring requirements of this section.
(a) The Department may invalidate a total coliform-positive sample only if the conditions of paragraph (3)(a)(i), (ii) or (iii) of this section are met. (i) The laboratory establishes that improper sample analysis caused the total coliform-positive result.(ii) The Department, on the basis of the results of repeat samples collected as required by paragraphs (2)(a) through (d) of this section, determines that the total coliform-positive sample resulted from a domestic or other non-distribution system plumbing problem. The Department cannot invalidate a sample on the basis of repeat sample results unless all repeat sample(s) collected at the same tap as the original total coliform-positive sample are also total coliform-positive, and all repeat samples collected within five service connections of the original tap are total coliform-negative (e.g., the Department cannot invalidate a total coliform-positive sample on the basis of repeat samples if all the repeat samples are total coliform-negative, or if the public water system has only one service connection).(iii) The Department has substantial grounds to believe that a total coliform-positive result is due to a circumstance or condition which does not reflect water quality in the distribution system. In this case, the system shall still collect all repeat samples required under paragraphs (2)(a) through (d) of this section, and use them to determine compliance with the MCL for total coliforms in Section F above. To invalidate a total coliform-positive sample under this paragraph, the decision with the rationale for the decision shall be documented in writing, and approved and signed by the supervisor of the Department official who recommended the decision. The Department shall make this document available to the EPA and the public. The written documentation shall state the specific cause of the total coliform-positive sample, and what action the system has taken, or will take, to correct this problem. The Department may not invalidate a total coliform-positive sample solely on the grounds that all repeat samples are total coliform-negative.(b) A laboratory shall invalidate a total coliform sample (unless total coliforms are detected) if the sample produces a turbid culture in the absence of gas production using an analytical method where gas formation is examined (e.g., the Multiple-Tube Fermentation Technique), produces a turbid culture in the absence of an acid reaction in the Presence-Absence (P-A) Coliform Test, or exhibits confluent growth or produces colonies too numerous to count with an analytical method using a membrane filter (e.g., Membrane Filter Technique). If a laboratory invalidates a sample because of such interference, the Department shall be notified, and the system shall collect another sample from the same location as the original sample within twenty-four (24) hours of being notified of the interference problem, and shall have it analyzed for the presence of total coliforms. The system shall continue to re-sample within twenty-four (24) hours and have the samples analyzed until it obtains a valid result. The Department may waive the twenty-four (24) hour time limit on a case-by-case basis.(4) Sanitary Surveys. (a)(i) Public water systems which do not collect five (5) or more routine samples per month shall undergo an initial sanitary survey by June 29, 1994, for community water systems and June 29, 1999, for non-community water systems. hereafter, systems shall undergo another sanitary survey every five (5) years, except that non-community water systems using only protected and disinfected ground water, as defined by the Department, shall undergo subsequent sanitary surveys at least every ten (10) years after the initial sanitary survey. The Department shall review the results of each sanitary survey to determine whether the existing monitoring frequency is adequate and what additional measures, if any, the system needs to undertake to improve drinking water quality.(ii) In conducting a sanitary survey of a system using ground water in a State having an EPA-approved wellhead protection program under the Federal Safe Drinking Water Act, information on sources of contamination within the delineated wellhead protection area that was collected in the course of developing and implementing the program should be considered instead of collecting new information, if the information was collected since the last time the system was subject to a sanitary survey.(b) Sanitary surveys shall be performed by the Department or an agent approved by the Department. The system is responsible for ensuring the survey takes place.(c) Sanitary surveys conducted by the Department under the provisions of 40 CFR 142.16(o)(2) may be used to meet the sanitary survey requirements of R.61-58.5.G(4).(5) Fecal coliforms/Escherichia coli (E. coli) testing. (a) If any routine or repeat sample is total coliform positive, the system shall analyze that total coliform-positive culture medium to determine if fecal coliforms are present, except that the system may test for E. coli in lieu of fecal coliforms. If fecal coliforms or E. coli are present, the system shall notify the Department by the end of the day when the system is notified of the test result, unless the system is notified of the result after the Department is closed, in which case the system shall notify the Department before the end of the next business day.(b) The Department has the discretion to allow a public water system, on a case-by-case basis, to forego fecal coliform or E. coli testing on a total coliform-positive sample if that system assumes that the total coliform-positive sample is fecal coliform-positive or E. coli-positive. Accordingly, the system shall notify the Department as specified in paragraph (5)(a) of this section and the provisions of Section F(2) above, apply.(6) Analytical methodology. (a) The standard sample volume required for total coliform analysis, regardless of analytical method used, is 100 ml.(b) Water systems need only determine the presence or absence of total coliforms; a determination of total coliform density is not required.(c) Analytical methods used to comply with R.61-58.5.G shall be in accordance with EPA-approved methods listed in 40 CFR 141 (11-8-06 edition).(d) Water systems must conduct fecal coliform analysis in accordance with the procedure outlined in 40 CFR 141.21(f)(5) (11-8-06 edition).(e) Water systems must conduct Escherichia coli analysis in accordance with the analytical methods outlined in 40 CFR 141.21(f)(6) (11-8-06 edition).(7) Response to violation. (a) A water system which has exceeded the MCL for total coliforms in Section F above, shall report the violation to the Department no later than the end of the next business day after it learns of the violation, and shall notify the public in accordance with R.61-58.6.E.(b) A water system which has failed to comply with a coliform monitoring requirement, including the sanitary survey requirement, shall report the monitoring violation to the Department within ten days after the system discovers the violation, and shall notify the public in accordance with R.61-58.6.E.(8) The provisions of R.61-58.5.G(1) and (4) are applicable until March 31, 2016. The provisions of R.61-58.5.G(2), (3), (5), (6), and (7) are applicable until all required repeat monitoring under R.61-58.5.G(2) and fecal coliform or E. coli testing under R.61-58.5.G(5) that was initiated by a total coliform-positive sample taken before April 1, 2016 is completed, as well as analytical method, reporting, recordkeeping, public notification, and consumer confidence report requirements associated with that monitoring and testing. Beginning April 1, 2016, the provisions of R.61-58.17 are applicable, with systems required to begin regular monitoring at the same frequency as the system-specific frequency required on March 31, 2016.S.C. Code Regs. § 61-58.5.G
Amended by State Register Volume 38, Issue No. 9, eff. 9/26/2014.