Current through Register Vol. 54, No.43, October 26, 2024
Section 223.12a - Fluoroscopic equipment(a) The fluoroscopic imaging assembly shall be provided with a primary protective barrier that intercepts the entire cross section of the primary beam at the maximum source to image receptor distance.(b) The X-ray tube used for fluoroscopy may not produce X-rays unless the primary barrier is in position to intercept the entire primary beam.(c) X-ray production in the fluoroscopic mode shall be controlled by a dead-man switch for the duration of any exposure. When recording serial fluoroscopic images, the fluoroscopist shall be able to terminate the X-ray exposures at any time. A means may be provided to permit completion of a single exposure of the series in process.(d) The protective barrier may not transmit more than 2 milliroentgens (.516 µC/kg) per hour at 10 centimeters from an accessible surface of the fluoroscopic imaging assembly for each roentgen per minute of entrance exposure rate.(e) During fluoroscopy and cinefluorography, the voltage and the current shall be continuously indicated.(f) A cumulative timing device activated by the fluoroscope switch shall be provided. It shall indicate the passage of a predetermined period of irradiation either by an audible signal or by temporary or permanent interruption of the irradiation when the increment of exposure time exceeds a predetermined limit not exceeding 5 minutes.(g) Fluoroscopic table designs when combined with normal operating procedures shall be of a type that no unprotected part of the staff or an ancillary individual's whole body is exposed to unattenuated scattered radiation which originates from under the table. The attenuation required may be not less than 0.25 millimeter lead equivalent.(h) Equipment configuration when combined with normal operating procedures shall be of a type that no portion of the staff or an ancillary individual's whole body, except the extremities, is exposed to the unattenuated scattered radiation emanating from above the tabletop unless one of the following criteria is met: (1) The individual is at least 120 centimeters from the center of the primary beam.(2) The radiation has passed through at least 0.25 millimeter of lead equivalent material-for example, drapes, bucky-slot cover (film-tray cover panel), sliding or folding panel or self-supporting curtains-in addition to the lead equivalency provided by the protective apron referred to in § 223.8(c) (relating to operating procedures).(i) In addition to the other requirements of this section, mobile fluoroscopes shall have image intensification.The provisions of this §223.12a adopted October 2, 1998, effective 10/3/1998, 28 Pa.B. 4894.The provisions of this §223.12a issued under sections 301 and 302 of the Radiation Protection Act (35 P. S. §§ 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).