25 Pa. Code § 221.41a

Current through Register Vol. 54, No.43, October 26, 2024
Section 221.41a - Fluoroscopic timer

A cumulative timing device activated by the fluoroscope switch shall be provided. It shall indicate the passage of a predetermined period of irradiation either by an audible signal or by temporary or permanent interruption of the irradiation when the increment of exposure time exceeds a predetermined limit not exceeding 5 minutes.