Or. Admin. Code § 333-122-0090

Current through Register Vol. 64, No. 1, January 1, 2025
Section 333-122-0090 - General Requirements for All RGD: Generator Cabinet

Each X-ray generator shall be supplied with a protective cabinet which limits leakage radiation measured at a distance of five centimeters from its surface such that it is not capable of producing a dose exceeding 0.25 millirem (2.5 µSv) in one hour.

Or. Admin. Code § 333-122-0090

PH 16-2021, adopt filed 04/21/2021, effective 4/21/2021

Statutory/Other Authority: ORS 453.605 - 453.807

Statutes/Other Implemented: ORS 453.605 - 453.807