Current through Register Vol. 64, No. 1, January 1, 2025
Section 333-122-0030 - General Requirements for All RGD: Warning Devices(1) Open-beam configurations shall be provided with a readily discernible indication of: (a) X-ray tube "on-off" status located near the radiation source housing, if the primary beam is controlled in this manner; and(b) Shutter "open-closed" status located near each port on the radiation source housing, if the primary beam is controlled in this manner. (2) Warning devices shall be labeled so that their purpose is easily identified. On equipment installed after January 1, 1978, warning devices shall have fail-safe characteristics. Or. Admin. Code § 333-122-0030
PH 16-2021, adopt filed 04/21/2021, effective 4/21/2021Statutory/Other Authority: ORS 453.605 - 453.807
Statutes/Other Implemented: ORS 453.605 - 453.807