Current through Register Vol. 63, No. 12, December 1, 2024
Section 333-121-0380 - Entering and Leaving the Radiation Room(1) Upon first entering the radiation room of a panoramic irradiator after an irradiation, the irradiator operator shall use a survey meter to determine that the source has returned to its fully shielded position. The operator shall check the functioning of the survey meter with a radiation check source prior to entry.(2) Before exiting from and locking the door to the radiation room of a panoramic irradiator prior to a planned irradiation, the irradiator operator shall: (a) Visually inspect the entire radiation room to verify that no one else is in it; and(b) Activate a control in the radiation room that permits the sources to be moved from the shielded position only if the door to the radiation room is locked within a preset time after setting the control.(3) During a power failure, the area around the pool of an underwater irradiator may not be entered without using an operable and calibrated radiation survey meter unless the over-the-pool monitor required by OAR 333-121-0140(2) is operating with backup power.Or. Admin. Code § 333-121-0380
PH 5-2005, f. & cert. ef. 4-11-05Stat. Auth.: ORS 453.675