The licensee must perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of:
(1) The source-specific input parameters required by the dose calculation algorithm;(2) The accuracy of dose, dwell time, and treatment time calculations at representative points;(3) The accuracy of isodose plots and graphic displays;(4) The accuracy of the software used to determine sealed source positions from radiographic images; and(5) The accuracy of electronic transfer of the treatment delivery parameters to the treatment delivery unit from the treatment planning system.Or. Admin. Code § 333-116-0605
PH 3-2003, f. & cert. ef. 3-27-03; PH 31-2004(Temp), f. & cert. ef. 10-8-04 thru 4-5-05; PH 36-2004, f. & cert. ef. 12-1-04; PH 12-2006, f. & cert. ef. 6-16-06; PH 4-2007, f. & cert. ef. 3-1-07; PH 59-2018, minor correction filed 02/01/2018, effective 2/1/2018Statutory/Other Authority: ORS 453.635
Statutes/Other Implemented: ORS 453.605 - 453.807