A licensee authorized to use radioactive material for imaging and localization studies must have in its possession a portable, radiation detection survey instrument capable of detecting dose rates over the range of one Sv (0.1 mrem) per hour to one mSv (100 mrem) per hour and a portable radiation measurement survey instrument capable of measuring dose rates over the range ten Sv (1 mrem) per hour to ten millisieverts (1000 mrem) per hour. The instruments must be operable and calibrated in accordance with OAR 333-116-0170.
Or. Admin. Code § 333-116-0350
Statutory/Other Authority: ORS 453.635
Statutes/Other Implemented: ORS 453.605 - 453.807