Or. Admin. Code § 333-106-0225

Current through Register Vol. 63, No. 11, November 1, 2024
Section 333-106-0225 - Fluoroscopic Source-to-Skin Distance

The source-to-skin distance shall not be less than:

(1) 38 cm on stationary fluoroscopes manufactured on or after August 1, 1974;
(2)35.5 cm on stationary fluoroscopes manufactured prior to August 1, 1974;
(3) 30 cm on all mobile fluoroscopes; and
(4) 20 cm for image intensified fluoroscopes used for specific surgical application. The written safety procedures must provide precautionary measures to be adhered to during the use of this device.

Or. Admin. Code § 333-106-0225

HD 4-1985, f. & ef. 3-20-85; HD 1-1991, f. & cert. ef. 1-8-91; PH 12-2006, f. & cert. ef. 6-16-06; PH 32-2014, f. 12-22-14, cert. ef. 1/1/2015

Stat. Auth.: ORS 453.605 - 453.807

Stats. Implemented: ORS 453.605 - 453.807