Okla. Admin. Code § 252:410-11-59

Current through Vol. 41, No. 19, June 17, 2024
Section 252:410-11-59 - Interruption and termination switches
(a)Interruption. The control panel of a system of 1 MeV or more must have a switch available to the operator during irradiation which allows the operator to interrupt irradiation and equipment movements at any time. Following an interruption, the operator must be able to restart irradiation by using a control panel switch without reselecting any operating conditions. However, if any pre-selected value is changed during an interruption, the system must prohibit further irradiation and equipment movements.
(b)Termination. Through switches at the control panel, an operator, at any time during irradiation, must be able to terminate irradiation and equipment movements or go from interruption status to complete termination.

Okla. Admin. Code § 252:410-11-59

Added at 17 Ok Reg 1136, eff 6-1-00