Okla. Admin. Code § 252:410-11-57

Current through Vol. 42, No. 7, December 16, 2024
Section 252:410-11-57 - Selection and display of dose monitor units

Therapeutic systems of 1 MeV or more shall be designed so that:

(1) irradiation cannot be possible until a selection of a number of dose monitor units has been made at the control panel;
(2) the pre-selected number of dose monitor units are displayed on the treatment control panel until reset manually for the next irradiation;
(3) after irradiation is terminated, the dosimeter display must be reset to zero before subsequent treatment can be initiated; and
(4) for post-1989 systems, after termination of irradiation, the preselected dose monitor units must be manually reset before irradiation can be initiated.

Okla. Admin. Code § 252:410-11-57

Added at 17 Ok Reg 1136, eff 6-1-00