Section 252:410-11-37 - Multiple tubes

When a control panel of a <1 MeV system can be used to energize multiple tubes, the operator must be able to activate only one x-ray tube at a time and confirm which x-ray tube is activated by viewing a control panel display. The machine must also have a display on each tube housing assembly which confirms that a specific tube is energized.