Current through all regulations passed and filed through October 28, 2024
Section 3701:1-52-26 - Entering and leaving the radiation room(A) Upon first entering the radiation room of a panoramic irradiator after an irradiation, the irradiator operator shall use a survey meter to determine that the source has returned to its fully shielded position. The operator shall check the functioning of the survey meter with a radiation check source prior to entry. (B) Before exiting from and locking the door to the radiation room of a panoramic irradiator prior to a planned irradiation, the irradiator operator shall: (1) Visually inspect the entire radiation room to verify that no one else is in it; and (2) Activate a control in the radiation room that permits the sources to be moved from the shielded position only if the door to the radiation room is locked within a preset time after setting the control. (C) During a power failure, the area around the pool of an underwater irradiator may not be entered without using an operable and calibrated radiation survey meter unless the over-the-pool monitor required by paragraph (B) of rule 3701:1-52-11 of the Administrative Code is operating with backup power. Ohio Admin. Code 3701:1-52-26