10A N.C. Admin. Code 41B.0322

Current through Register Vol. 39, No. 9, November 1, 2024
Section 41B .0322 - INTOXIMETERS: MODEL INTOX EC/IR II AND MODEL INTOX EC/IR II (ENHANCED WITH SERIAL NUMBER 10,000 OR HIGHER)

The following operational procedures shall be followed when administering a test using the Intoximeters, Model Intox EC/IR II and Model Intox EC/IR II (Enhanced with serial number 10,000 or higher):

(1) Ensure instrument displays time and date;
(2) Ensure observation period requirements have been met;
(3) Initiate breath test sequence;
(4) Enter information as prompted by the instrument;
(5) Verify instrument accuracy;
(6) When "PLEASE BLOW" appears, collect breath sample;
(7) When "PLEASE BLOW" appears, collect breath sample; and
(8) Print test record.

If the alcohol concentrations differ by more than 0.02, a third or fourth breath sample shall be collected when "PLEASE BLOW" appears. Sequential tests shall be administered as soon as feasible by repeating steps (1) through (8), as applicable.

10A N.C. Admin. Code 41B .0322

Authority G.S. 15A-534.2(d); 20-139.1(b);
Eff. November 1, 2007;
Pursuant to G.S. 150B-21.3A, rule is necessary without substantive public interest Eff. January 9, 2018;
Amended Eff. April 1, 2020.
G.S. 20-139.1(b);
Eff. November 1, 2007.
Amended by North Carolina Register Volume 34, Issue 21, May 1, 2020 effective 4/1/2020.