Section 8:27-5.3 - Sterilization procedures(a) All instruments that are processed by steam sterilization must first be cleaned. The manufacturer's instructions of the autoclave regarding water purity requirements, filling, draining, and general maintenance shall be followed. A copy of the instructions shall be maintained on site.(b) Peel-packs shall be positioned standing on edge, paper to plastic. Loading racks or baskets specifically designed for these types of packages, or other means of holding them on edge and properly spaced, shall be used.(c) The manufacturer's written instructions of the autoclave for the cycle parameters, time, temperature and pressure shall be followed.(d) Policies and procedures shall be established when the cycle does not include a drying phase. Drying cycle shall be in accordance with the manufacturer's instructions.(e) Wrapped items being cooled after removal from the autoclave shall remain untouched in the loading tray during the cooling period.(f) All hinged instruments shall be processed in an open position.N.J. Admin. Code § 8:27-5.3