TABLE 4 | ||
Medical Physicist's Radiographic QC Survey | ||
Item | Test | Standard |
1. | Radiographic Unit Assembly | As required at 7:28-15.3 |
Evaluation | ||
2. | Collimation Assessment | As required at 7:28-15.3 |
3. | Collimator Illumination | As required at 7:28-15.3 |
4. | Half Value Layer | As required at 7:28-15.3 |
5. | mA Exposure Linearity | As required at 7:28-15.3 |
6. | kVp Accuracy/Reproducibility | As required at 7:28-15.3 |
7. | Timer Accuracy/Reproducibility | As required at 7:28-15.3 |
8. | Automatic Exposure Control, | As required at 7:28-15.3 |
Reproducibility, Tracking, | ||
Density Control | ||
9. | Entrance Skin Exposure (ESE) | Determine ESE for common exam and compare |
Measurement | with National Evaluation of X-ray Trends | |
(NEXT) data available in the Compliance | ||
Guidance Documents referenced at N.J.A.C. | ||
7:28-22.3(c)2 | ||
10. | Image Quality Evaluation | Established standard for phantom test tool |
(Recommendation) | used | |
11. | Review Facility/Technologist | Review QC tests for proper procedure and |
QC Test Records | corrective action | |
12. | Physicist Report and | Communicate results and recommendations to |
Recommendations | registrant |
N.J. Admin. Code § 7:28-22.8