TABLE 2 | ||||
Fluoroscopic Quality Control Requirements | ||||
(To be performed by appropriately trained facility personnel) | ||||
Item | Required Test or | Frequency | Standard | |
Procedure | ||||
1. | Equipment Warm-up | Daily, each | Tube warm-up and ensure equipment is | |
Procedure | day fluorosco | working properlyFluoro phantom image is | ||
py is perform | comparable to facility standard | |||
ed | ||||
2. | Laser Film Printer | Weekly | Recommended control | *OD=Optical |
Quality Control | limits | Density | ||
SMPTE Test Pattern | Inverted gray | |||
scale | ||||
0% patch 2.45 +/- | 0% patch 2.50 + | |||
0.15 OD* | 0.15 OD | |||
10% patch 2.10 +/- | 10% patch 2.25 +/- | |||
0.15 OD | 0.15 OD | |||
40% patch 1.15 +/- | 40% patch 1.35 +/- | |||
0.15 OD | 0.15 OD | |||
90% patch 0.30 +/- | 90% patch 0.30 +/- | |||
0.08 OD | 0.08 OD | |||
*The 5% patch should just be visible | ||||
inside of the 0% patch. | ||||
The 95% patch should be visible inside | ||||
the 100% patch. | ||||
3. | For spot film and | As specified | As specified in N.J.A.C. 7:28-22. 5 | |
radiography, items | in Table 1, | Table 1, Radiographic Quality Control | ||
2, 4, 5, 7, 9 and11 | Radiographic | Requirements | ||
QC tests as | Quality | |||
specified in Table | Control | |||
1, Radiographic | Requirements | |||
Quality Control | ||||
Requirements | ||||
4. | Phantom Images | Monthly | kVp +/- 5%, MA +/- 10% high & low | |
(Fluoro Video | contrast depends on phantom used | |||
Monitor) | ||||
5. | Equipment Visual | Initially and | All tests passed | |
Checklist | quarterly | |||
6. | Lead Aprons, | Initially and | No breaks in protective garments | |
Gloves, Gonadal and | annually | |||
Thyroid Shield | ||||
Integrity Check | ||||
7. | Medical Physicist's | Initially and | As required in 7:28-22.9 | |
QC Survey | annually | |||
8. | Quality Assurance | Initially and | As required in 7:28-22.4(a)7 | |
Program Review | annually |
N.J. Admin. Code § 7:28-22.6