N.H. Admin. Code § He-P 4046.09

Current through Register No. 45, November 7, 2024
Section He-P 4046.09 - Hand-held Dental Intraoral Machines

The following requirements shall apply to the use of hand-held dental intraoral machines:

(a) For all uses:
(1) The registrant shall follow all applicable requirements set forth in the New Hampshire Rules for the Control of Radiation (NHRCR), as well as all manufacturers's requirements;
(2) The machine shall be used only for dental diagnostic imaging;
(3) The protective shielding shall not be removed or modified;
(4) The registrant shall provide a copy of their operating, safety and security procedures to operators to prevent unauthorized or improper use;
(5) The registrant shall prepare a quality assurance plan, with an emphasis on ensuring limited retaking of radiographs;
(6) Operators of hand-held dental intraoral machines shall be specifically trained to operate such equipment, and the registrant shall provide proof of training for all operators;
(7) When operating a hand-held dental intraoral machine, operators shall wear a lead apron as well as an extremity dosimeter on the hand closest to the protective shielding;
(8) A hand-held dental intraoral machine shall be held without any motion during a patient examination. A tube stand or Rinn kit may be utilized to immobilize a hand-held dental intraoral radiographic unit during patient examination;
(9) Operators of hand-held dental intraoral machines shall provide a secondary protective barrier for patients; and
(10) The operator shall ensure that there are no bystanders within a radius of at least 6 feet (2 meters), from the patient being examined with a hand-held intraoral machine.
(b) Additional requirements for operatories in permanent facilities:
(1) Hand-held dental intraoral machines shall be used for patient examinations only in dental operatories that meet the structural shielding requirements specified by DHHS/RHS or by a qualified expert;
(2) Hand-held dental intraoral machines shall not be used for patient examinations in hallways and waiting rooms; and
(3) For battery-powered units, visual means shall be provided on the control panel to indicate whether the battery is in a state of charge adequate for proper operation.