The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing shall include, as applicable, verification of:
(a) The source-specific input parameters required by the dose calculation algorithm;(b) The accuracy of dose, dwell time, and treatment time calculations at representative points;(c) The accuracy of isodose plots and graphic displays;(d) The accuracy of the software used to determine sealed source positions from radiographic images; and(e) The accuracy of electronic transfer of the treatment delivery parameters to the treatment delivery unit from the treatment planning system.N.H. Admin. Code § He-P 4035.57
(See Revision Note at part heading for He-P 4035) #6942, eff 2-1-99; ss by #8800, INTERIM, eff 2-1-07, EXPIRED: 7-31-07
New. #8959, eff 8-7-07
Amended by Volume XXXV, Number 32 Filed August 13, 2015, Proposed by #10897, Effective 7/22/2015, Expires1/18/2016.Amended by Volume XXXVI Number 01, Filed January 7, 2016, Proposed by #11011, Effective 1/15/2016, Expires 1/15/2026.The amended version of this section by New Hampshire Register Volume 36, Number 49, eff.11/19/2016 is not yet available.