Current through Register No. 45, November 7, 2024
Section He-P 4035.52 - Technical Requirements for Mobile Remote Afterloader(a) A licensee providing mobile remote afterloader service shall: (1) Check survey instruments before medical use at each address of use or on each day of use, whichever is more frequent; and(2) Account for all sources before departure from a client's address of use.(b) In addition to the periodic spot-checks required by He-P 4035.55(i) through (o), a licensee authorized to use mobile afterloaders for medical use shall perform checks on each remote afterloader unit before use at each address of use. At a minimum, checks shall be made to verify the operation of: (1) Electrical interlocks on treatment area access points;(2) Source exposure indicator lights on the remote afterloader unit, on the control console, and in the facility;(3) Viewing and intercom systems;(4) Applicators, source transfer tubes, and transfer tube-applicator interfaces;(5) Radiation monitors used to indicate room exposures;(6) Source positioning (accuracy); and(7) Radiation monitors used to indicate whether the source has returned to a safe shielded position.(c) In addition to the requirements for checks in He-P 4035.52(b), a licensee shall ensure overall proper operation of the remote afterloader unit by conducting a simulated cycle of treatment before use at each address of use.(d) If the results of the checks required in He-P 4035.52(b) indicate the malfunction of any system, a licensee shall lock the control console in the "off" position and not use the unit except as may be necessary to repair, replace, or check the malfunctioning system.(e) A licensee shall retain a record of each check required by He-P 4035.52(b) for 3 years. The record shall include: (1) The date of the check;(2) The manufacturer's name, model number, and serial number of the remote afterloader unit;(3) Notations accounting for all sources before the licensee departs from a facility;(4) Notations indicating the operability of each entrance door electrical interlock, radiation monitors source exposure indicator lights, viewing and intercom system, applicators source transfer tubes, and transfer tube applicator interfaces, and source positioning accuracy; and(5) The name and the signature of the individual who performed the check.N.H. Admin. Code § He-P 4035.52
(See Revision Note at part heading for He-P 4035) #6942, eff 2-1-99; ss by #8800, INTERIM, eff 2-1-07, EXPIRED: 7-31-07
New. #8959, eff 8-7-07
Amended by Volume XXXV, Number 32 Filed August 13, 2015, Proposed by #10897, Effective 7/22/2015, Expires1/18/2016.Amended by Volume XXXVI Number 01, Filed January 7, 2016, Proposed by #11011, Effective 1/15/2016, Expires 1/15/2026.Amended by Volume XL Number 24, Filed June 11, 2020, Proposed by #13048, Effective 6/2/2020, Expires 6/2/2030.