Current through November 8, 2024
Section 484C.120 - Calibration of device used by law enforcement agency1. Each evidential breath-testing device used by a law enforcement agency must be calibrated by a certified forensic analyst of alcohol at least once within the 90 days immediately preceding the date on which the device is used to test a person's breath.2. The calibration must include verification by the analyst of the response and accuracy of the device at no less than three levels of alcohol within the range which corresponds to a concentration of alcohol in the breath from 0 to 0.4 gram per 210 liters of breath, inclusive.Nev. Admin. Code § 484C.120
(Added to NAC by Com. on Testing for Intoxication, eff. 10-26-83; A 3-11-86; 9-9-94; R137-99, 12-13-99; R032-05, 10-31-2005) - (Substituted in revision for NAC 484.660)