Section 459.581 - Registration of electronic brachytherapy systems or additional therapeutic X-ray systems1. Except as otherwise provided in subsection 4, each person who acquires an electronic brachytherapy system or an additional therapeutic X-ray system shall apply to the Division for registration of the machine within 30 days after installing the machine. The application must include, without limitation: (a) A list of all authorized users, radiation therapy physicists and operators;(b) The name of the radiation safety officer and radiation safety committee members;(c) A copy of the most recent record of surveys, calculations and quality assurance checks on each machine;(d) A current copy of the quality management program created pursuant to NAC 459.559;(e) A current copy of the quality assurance program created pursuant to NAC 459.5937; and(f) The manufacturer's certification.2. No medical therapy device may be used on a person until the facility has received a certificate of registration from the Division.3. A separate registration is required for facilities which:(b) Are not under a single radiation safety program; or(c) Are not under the same management.4. The provisions of this section do not apply to radiation devices which are in transit or in storage.Nev. Admin. Code § 459.581
Added to NAC by Bd. of Health by R185-08, eff. 5-7-2010