Section 444.00911 - Requirements for sterilizer; check of expiration date before use
1. A sterilizer used in an invasive body decoration establishment must:
(a) Be manufactured for the sterilization of medical instruments;
(b) Meet the specifications set forth in the standards adopted by reference in paragraph (a) of subsection 1 of section 35;
(c) Be tested using a commercial biological indicator monitoring system:
(1) After the initial installation;
(2) After any service or repair; and
(3) Once each month; and
(d) Be loaded, operated, disinfected and maintained according to the manufacturer's specifications
2. Before each use, the operator of the sterilizer must check the expiration date of the biological indicator monitoring system.