180 Neb. Admin. Code, ch. 7, § 074

Current through September 17, 2024
Section 180-7-074 - FULL CALIBRATION MEASUREMENTS ON REMOTE AFTERLOADER UNITS
074.01FREQUENCY. A licensee authorized to use a remote afterloader unit for medical use must perform full calibration measurements on each unit:
(A) Before the first medical use of the unit;
(B) Before medical use under the following conditions:
(i) Following replacement of the source or following reinstallation of the unit in a new location outside the facility; and
(ii) Following any repair of the unit that includes removal of the source or major repair of the components associated with the source exposure assembly;
(C) At intervals not exceeding one calendar quarter for HDR, MDR, and PDR remote afterloader units with sources whose half-life exceeds 75 days; and
(D) At intervals not exceeding one year for LDR remote afterloader units.
074.02DETERMINANTS. To satisfy the requirement of 7-074.01, full calibration measurements must include, as applicable, determination of:
(A) The output within ± 5%;
(B) Source positioning accuracy to within ±1 millimeter;
(C) Source retraction with backup battery upon power failure;
(D) Length of the source transfer tubes;
(E) Timer accuracy and linearity over the typical range of use;
(F) Length of the applicators; and
(G) Function of the source transfer tubes, applicators, and transfer tube-applicator interfaces.
074.03VERIFICATION OF INVENTORY AND SOURCE ARRANGEMENT. In addition to the requirements for full calibration for LDR remote afterloader units in 180 NAC 7-074.02, a licensee must perform an autoradiograph of the source or sources to verify inventory and source or sources arrangement at intervals not exceeding one calendar quarter.
074.04USE OF DOSIMETRY SYSTEM. A licensee must use the dosimetry system described in 180 NAC 7-072.01 to measure the output.
074.05PROTOCOLS. A licensee must make full calibration measurements required by 180 NAC 7-074.01 according to published protocols accepted by nationally recognized bodies.
074.06SOURCE MANUFACTURER MEASUREMENTS. For LDR remote afterloader units, a licensee may use measurements provided by the source manufacturer that are made according to 180 NAC 7-074.01 through 7-074.05.
074.07ONE PERCENT PHYSICAL DECAY. A licensee must mathematically correct the outputs determined in 180 NAC 7-074.02(A) for physical decay at intervals consistent with one percent physical decay.
074.08FULL CALIBRATION MEASUREMENTS AND PHYSICAL DECAY CORRECTIONS. Full calibration measurements required by 180 NAC 7-074.01 and physical decay corrections required by 180 NAC 7-074.07 must be performed by the authorized medical physicist.
074.09RECORDS. A licensee must retain a record of each calibration according to 180 NAC 7-108.

180 Neb. Admin. Code, ch. 7, § 074

Amended effective 11/3/2020
Amended effective 6/14/2023