The licensee must perform acceptance testing on the treatment planning system of therapy-related computer systems according to published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of:
(A) The source-specific input parameters required by the dose calculation algorithm;(B) The accuracy of dose, dwell time, and treatment time calculations at representative points;(C) The accuracy of isodose plots and graphic displays; and(D) The accuracy of the software used to determine sealed source positions from radiographic images.180 Neb. Admin. Code, ch. 7, § 061